        <p>目的 应用 ATP 生物荧光法评价两种不同清洗方法对手术器械的清洗效果。方法 随机选择使用后的手术器械为研究对象，并分为研
究组和对照组，各组 110 件。研究组使用超声波预洗和全自动清洗消毒机清洗，对照组则采用超声波预洗和手工清洗。通过对比两组的清
洗质量以及组内常规器械和管腔器械的清洗效果，来评估不同清洗方法的优劣。结果 通过裸眼/有光源放大镜及 ATP 生物荧光测定两种方
法对器械清洗质量进行对比，结果显示研究组器械清洗合格率显著高于对照组。通过裸眼/有光源放大镜观察到，研究组和对照组器械清洗
合格率分别为 96.36%和 100.00%，有显著性差异(X²=4.060，P</p>