        <article-title>血细胞分析仪测血小板结果偏低的原因分析及其纠正方法</article-title>
      </title-group>
      <contrib-group>
        <contrib contrib-type="author">
          <string-name>窦鹏 (内蒙古包头市昆都仑区医院 014010）</string-name>
        </contrib>
      </contrib-group>
      <pub-date pub-type="epub">
        <year>2024</year>
        <month>18</month>
      </pub-date>
      <issue>18</issue>
      <abstract>
        <p>目的：探讨血细胞分析仪测血小板结果偏低的原因分析及其纠正方法。方法：选取2023年9月~2024年7月收治的住院患者血小板 结果偏低的血样样本360例，分析引起血小板结果偏低的原因与纠正方法。结果：血细胞分析仪检测血小板结果偏低的主要原因为抗凝剂 因素（25.56%）、血液标本不合格因素（25.28%）、血小板体积偏大（25.00%）与病理因素（24.17%）；血细胞分析仪检测的血小板结果明 显低于显微镜计数结果其结果差异存在统计学意义（P＜0.05）；纠正后的血细胞分析仪的血小板计数结果与显微镜计数相比无统计学意义 （P＞0.05）。结论：血细胞分析仪检测血小板计数偏低的原因相对较多，需要及时采取针对性纠正方法，以确保血小板计数结果的准确性， 避免偏低结果的出现。</p>